Synopsis
The Nihon Kohden JE-120A amplifier is a clinical monitoring system capable of acquiring EEG or ECoG signals from up to 256 channels. Paired with the QI-123A LAN converter box, this system can provide a research data stream in addition to the clinical data stream. The research data stream can be sampled at up to 2kHz for 256 channels, 5kHz for 128 channels or 10kHz for 64 channels.
Location
http://www.bci2000.org/svn/trunk/src/private/SignalSource/NihonKohdenSource
Versioning
Authors
Griffin Milsap (griffin.milsap@gmail.com) Johns Hopkins University
Peter Brunner (brunner@neurotechcenter.org) Wadsworth Center - National Center for Adaptive Neurotechnologies
Version History
Version 1.0: November 2015
- #
Source Code Revisions
- Initial development: 4913
- Tested under: #
- Known to compile under: 4913
- Broken since: --
Functional Description
Equipment Setup
Installation
Parameters
- SourceCh total number of channels to be acquired from the source module or auto for all channels (256 channels)
- SampleBlockSize number of samples acquired in each block or auto for default setting (1/10th of sampling rate)
- NOTE: Only specific combinations of SampleBlockSize and SampleRate are valid! See #Valid Operating Modes
- SampleRate determines the rate at which the device samples data.
- NOTE: Only specific combinations of SampleBlockSize and SampleRate are valid! See #Valid Operating Modes
- ChannelNames is a convenience parameter. Name channels here and they can be referenced by these names later.
- SourceChOffset auto for default settings (0 for each channel).
- SourceChGain auto for default settings (1 for each channel).
- SourceChList list of channels that should be acquired or auto for all channels
- SourceBufferSize
Valid Operating Modes
The NihonKohdenSource only accepts specific sampling rates. The number of channels acquire is limited at sampling rates of 5000 Hz and above. The following table shows all valid combinations of sample rate and maximum number of channels.
SamplingRate | Max SourceCh | Aliasing Filter Cut-Off |
---|---|---|
[samples/sec] | [channels] | [Hz] |
100 | 256 | 30 or 35 |
200 | 256 | 60 or 70 |
500 | 256 | 120 |
1000 | 256 | 300 |
2000 | 256 | 600 |
5000 | 128 | 1200 |
10000 | 64 | 3000 |
Valid Operating Modes
State Variables
- DC01 to DC16 DC Input data
- Digital Range: 0 to 65535
- Analog Range: 0xFFFF(Max)-0x8000(0V)-0x0000(Min). (Signed.)
- 1LSB: 366.3uV
- SpO2
- Digital Range: 0 to 100
- 0 means 0%, 100 means 100%
- EtCO2
- Digital Range: 0 to 100
- 0 means 0mmHg, 100 means 100mmHg
- BP1 to BP2
- not implemented
- DigitalMark1 to DigitalMark2
- not implemented